Test Code RFIGA Rheumatoid Factor IgA
Additional Codes
Meditech Order: RFIGA
Panel Order:RAPANEL
Intended Use
EliA RF IgA is intended for the in vitro quantitative measurement of IgA rheumatoid factor in serum and plasma (Li-heparin, EDTA) to aid in the diagnosis of rheumatoid arthritis. EliA RF IgA uses the EliA IgA method on the Phadia 250 instrument.
Performing Lab
Boone Health Laboratory
Collection Method
Venous Collection
Container
Gold Serum Tube
Instructions/Special Handling
None
Preferred Specimen
Gold SST Serum Tube
Red Serum
Acceptable Specimes
- Plasma
- Lithium Heparin
- EDTA
Units
| EliA U/mL |
Reference Ranges
| <14 | Negative |
| >20 | Positive |
| 14-20 | Equivoval (recommend retesting after 8-12 weeks) |
| >214 | >214 EliA U/mL |
Critical Values
N/A
Reasons for Rejection
| Hemolysis | no hemolysis |
| Icterus | |
| Lipemia | no lipemia |
| Other | Contaminated specimen, QNS, Improperly labeled, Improper specimen type |
Specimen Stability
| Ambient (15-25°C) | 8 hours |
| Refrigerated (2-8°C) | 14 days |
| Frozen (-20°C) | 12months (only one freeze thaw cycle) |
Performance Information
| Days/Times Performed | Monday, Wednesday and Friday |
| Expected Turn around Time | 1-3 days |
| Methodology/Method Description | EliA IgA |
CPT Code(s)
86431